Jenway 62 Fluorimeter Series

Three Jenway fluorimeter models are available to cover a wide range of applications. The Fluorimeter 6280 model is ideal for the most sensitive determinations with emission wavelengths up to 650nm. When higher emission wavelengths are required, the Jenway 6285 with its red-enhanced detector is the Fluorimet of choice. For less sensitive applications and a broader wavelength range up to 1100nm the low cost 6270 unit is more suitable.

All models are powered from an external, universal mains adapter suitable for use from 90V to 230V, while the 12V DC input enables use in vehicles or from suitable battery packs.
 

 
Excitation and emission filters are not included and should be purchased as separate items. Please refer to the filter section. Click Here To Order Fluorimeter Filters
 
• High quality optics: pulsed xenon lamp and Total Energy Transfer (TET)
• Intelligent Filter Modules ensure correct filters are fitted for selected method
• Intuitive programming allows fast program creation
• Raw fluorescence or concentration values options
• Standard curve and kinetic functions
• Heated sample holder and sipper pump accessories
• Compatible with DataWay PC software

 
Technical Specification
 

Parameter Jenway Fluorimeter 6270 Jenway Fluorimeter 6280 Jenway Fluorimeter 6285
Wavelength range 190 to 1100nm 190 to 650nm 190 to 850nm
Sensitivity1 1µg/ml <1ng/ml <1ng/ml
Dynamic range1 5.2x104 5.2x106 5.2x106
Detector Photodiode PMT2

PMT2

 
1 Using quinine sulphate 2 Photomultiplier tube

SKU# 627 001

Model 6270 Fluorimeter (190-1100nm) supplied with 12V universal power supply with UK, EU and US mains leads, pack 100 disposable cuvettes and operating instructions.

 

Excitation and emission filters are not included and should be purchased as separate items. Please refer to the filter section. Please Click Here To Order

Technical Specification


 

Parameter 6270
Wavelength range 190 to 1100nm
Sensitivity1 1µg/ml
Dynamic range1 5.2x104
Detector Photodiode
 
1 Using quinine sulphate
2 Photomultiplier tube
 
Features

Feature 62 Series Fluorimeter
Wavelength selec tion Intelligent Filter Modules (IFM)
Light source Pulsed Xenon lamp with press-to-read operation
Expected life of lamp No replacement required in normal operation
Raw fluorescence readings Yes
Concentration standard curve Blank + up to 6 standards
Kinetics, timed option In both raw fluorescence and concentration modes
Kinetics options 1 reading every 3 to 999 seconds
No. methods stored 20
No. results stored 100 per method
GLP Method lock with password access. Time and date stamped results. Last calibration date and time stored. Calibration interval set. Calibration due reminder. User ID
Outputs RS232
PC software Compatible with DataWay (option)
Sample handling 10mm square cuvette holder (standard). Heated sample holder (option). Sipper pump (option)
Power supply 90 to 230V universal mains adapter 12V DC input
Size (w x d x h) 365 x 272 x 160mm
Weight 6kg

 


Availability: IN STOCK
SKU# 628 001

Model 6280 Fluorimeter (190-650nm) supplied with universal 12V power supply with UK, EU and US mains leads, pack 100 disposable cuvettes and operating instructions.

 

Excitation and emission filters are not included and should be purchased as separate items. Please refer to the filter section. Please Click Here To Order

Technical Specification


 
Parameter 6280
Wavelength range 190 to 650nm
Sensitivity1 <1ng/ml
Dynamic range1 5.2x106
Detector PMT2


1 Using quinine sulphate
2 Photomultiplier tube

 Features


Feature 62 Series Fluorimeter
Wavelength selec tion Intelligent Filter Modules (IFM)
Light source Pulsed Xenon lamp with press-to-read operation
Expected life of lamp No replacement required in normal operation
Raw fluorescence readings Yes
Concentration standard curve Blank + up to 6 standards
Kinetics, timed option In both raw fluorescence and concentration modes
Kinetics options 1 reading every 3 to 999 seconds
No. methods stored 20
No. results stored 100 per method
GLP Method lock with password access. Time and date stamped results. Last calibration date and time stored. Calibration interval set. Calibration due reminder. User ID
Outputs RS232
PC software Compatible with DataWay (option)
Sample handling 10mm square cuvette holder (standard). Heated sample holder (option). Sipper pump (option)
Power supply 90 to 230V universal mains adapter 12V DC input
Size (w x d x h) 365 x 272 x 160mm
Weight 6kg

 



Availability: IN STOCK
SKU# 628 501

Model 6285 Fluorimeter (190-850nm) supplied with 12V universal power supply with UK, EU and US mains leads, pack 100 disposable cuvettes and operating instructions.

 

Excitation and emission filters are not included and should be purchased as separate items. Please refer to the filter section. Please Click Here To Order

Technical Specification

 

Parameter 6285
Wavelength range 190 to 850nm
Sensitivity1 <1ng/ml
Dynamic range1 5.2x106
Detector PMT2
 
1 Using quinine sulphate
2 Photomultiplier tube

 
Features

Feature 62 Series Fluorimeter
Wavelength selec tion Intelligent Filter Modules (IFM)
Light source Pulsed Xenon lamp with press-to-read operation
Expected life of lamp No replacement required in normal operation
Raw fluorescence readings Yes
Concentration standard curve Blank + up to 6 standards
Kinetics, timed option In both raw fluorescence and concentration modes
Kinetics options 1 reading every 3 to 999 seconds
No. methods stored 20
No. results stored 100 per method
GLP Method lock with password access. Time and date stamped results. Last calibration date and time stored. Calibration interval set. Calibration due reminder. User ID
Outputs RS232
PC software Compatible with DataWay (option)
Sample handling 10mm square cuvette holder (standard). Heated sample holder (option). Sipper pump (option)
Power supply 90 to 230V universal mains adapter 12V DC input
Size (w x d x h) 365 x 272 x 160mm
Weight 6kg

 